41
IMPORTANT
• Was the sample free of clear or watery uid?
• Was the drop large enough (at least 0.3 µL)?
• Did the reaction site ll with enough blood all at once?
Step 5:
The meter will count down, with a ashing dash, and display
result in 7 seconds.
CONSIDER YOUR RESULT. REPEAT ALTERNATE SITE TEST:
• IF the blood sample appeared to be diluted with clear uid.
• IF you did not vigorously rub the test site.
• IF the blood drop was not large enough to ll the
reaction site.
• IF the test was accidentally marked as a “Control” result.
• IF your result was not consistent with how you feel.
• IF more than 20 seconds elapsed from sample
collection to measurement (evaporation of the
blood sample may cause a test result that is higher
than the actual value).
IF NOT,
REPEAT TEST.
Filled
Not Filled
Any of the above situations can lead to an inaccurate test result.