It is Recommended to Perform a Control Solution Test:
Whenever you suspect the meter or test strips may not be
functioning properly.
If test results appear to be abnormally high or low or are not
consistent with clinical symptoms.
The test strip bottle has been left open or has been exposed to
temperatures below 34°F (1°C) or above 86°F (30°C), or humidity
levels above 80%.
To check your technique.
When the ReliOn
®
Conrm blood glucose meter has been
dropped or stored below 32°F (0°C) or above 122°F (50°C).
When you open a new bottle of test strips.
Important:
Use only with ReliOn
®
Conrm blood glucose meter and
ReliOn
®
Conrm/Micro blood glucose test strips.